Brand:
GESTERItem NO.:
GT-RA12-3Order(MOQ):
1Product Detail
Application: In order to prevent infection at the medical site and during labor work, workers must be protected from inhalable hazardous substances in the workplace, and respirators such as masks must be worn during work. Select a respirator, such as a mask, based on your facial features, and evaluate the tightness between the respirator, such as a mask and the face, to check whether there are gaps or leaks that put workers at risk. The mask tightness tester can quickly complete the tightness test of respirators such as masks to ensure that it provides good protective performance. Security experts will also develop protection schemes and standard regulations based on the results of the tightness test. Widely used in hospitals, manufacturing plants, production sites, fire-fighting workplaces and other official testing agencies. Standards:
GB 19083-2010 Technical Requirements for Medical Protective Masks GB2626-2019 Respiratory protection—Non-powered air-purifying particle respirator
Features:
|
How to choose a high-quality mask not only depends on the filtering efficiency of the mask, but also to confirm whether the mask is completely in contact with the face, otherwise aerosol particles such as germs that have not been filtered by the mask will be inhaled from the inadequate adhesion.
Therefore, it is very important to check the wearing status of the mask. The European and American countries have listed the application test of the mask as a strong inspection item.
For example, the N95 mask is one of 9 types of particulate protective masks certified by NIOSH (National Institute of Occupational Safety and Health). "N" indicates oil resistance. "95" indicates that the particle concentration in the mask is more than 95% lower than that of the particles outside the mask when exposed to the specified number of special test particles. N95 is not a specific product name, as long as it meets the N95 standard and passes the NIOSH review, it can be called "N95 mask".
The mask tightness tester can quickly complete the tightness test of respirators such as masks to ensure that it provides good protection performance.
Model |
GT-RA12-3 |
Concentration range |
0 ~ 100,000 / cm3 |
Grain diameter |
0.02 ~ 1.0μm |
Flow |
Sampling flow: 100cm3 / min Total flow: 700cm3 / min |
Closeness coefficient test |
direct test (Cout / Cin) |
Liquor |
99.5% + isopropanol (analytical grade) |
Display |
7inch true color touch screen |
Communication interface |
USB × 3 (Host × 2, Device × 1) Ethernet interface × 1 |
Connection port |
environment port, sampling port |
WIFI |
equipped |
Languages |
English, French, Spanish, Portuguese, Chinese |
Flow control |
sensor control |
PC controllable operation |
one computer can control 4 instruments at the same time |
Data output format |
Microsoft Excel |
Working temperature |
15 ~ 35 ℃ |
Power supply |
AC 110 ~ 240V 50 / 60Hz |
Appearance size |
208×117×262mm |
Weight |
2.1kg |
Accessories:
Standard Accessories |
Alcohol reagent bottle |
Protective cap |
|
Reagent stick |
|
Zero-count filter |
|
Strainer |
|
Sampling tube |
|
Instruction manual |
|
AC adapter |
|
Touch screen pen |
|
Optional accessories |
Computer |
Test kit for tightness coefficient |