Brand:
GESTERItem NO.:
GT-RA02BOrder(MOQ):
1Product Detail
Application: Mask Bacteria Filtration Efficiency Tester,are mainly used for testing the bacterial filtration efficiency performance of masks applicable to metrological verification departments, scientific research institutes, medical mask production units and their related testing departments. It has the characteristics of fast and accurate measurement of the bacterial filtration efficiency of masks [BFE].
The Mask Bacterial Filtration Efficiency (BFE) Detector is composed of a biological aerosol generating system, aerosol chamber, aerosol delivery device, negative pressure cabinet, and 28.3LPM Anderson sampler. The entire detector is controlled by the control system. The console uses PLC+Labview to coordinate and control the work of the aerosol generation system, transmission system, negative pressure cabinet, and sampling system, and displays the working status in real time, and the entire measurement work is automatically completed. Standards:
YY 0469 Technical requirements for surgical mask |
9.The door of the test chamber is equipped with large observation window and operation gloves, so as to observe during the test and operate the test.
Model |
GT-RA02B |
A Route Sampling Flow |
28.3L/min, 0.01L/min |
B Route Sampling Flow |
28.3L/min, 0.01L/min |
Pressure of A Route Sampling Flowmeter |
(-5000~0)Pa |
Pressure of B Route Sampling Flowmeter |
(-5000~0)Pa |
Pressure of Spray Flowmeter |
(0~300)kPa |
Vacuum Pump |
the maximum flow rate :57LPM; Equipped with silencer, stable air flow, negative pressure: -900pa |
Peristaltic Pump Flow |
(0.01~3.0)mL/min |
Working Temperature |
0-50℃ |
Negative Pressure of Aerosol chamber |
Pressure sensor range (-5000~0)Pa, negative pressure can be adjusted, normal operation negative pressure (-500~0)Pa |
Negative Pressure of Main Chamber
|
Pressure sensor range:(-500~0)Pa, negative pressure can be adjusted by fan speed, |
Data Save Ability |
Data can be downloaded via U disk with a storage capacity of up to 8G |
High Efficiency Particulate Air Filter Properties |
≥99.99%@0.3μm, |
Median diameter of aerosol generator mass |
Average diameter: (3.0±0.3)μm; Geometric Standard Deviation≤1.5 |
Double Routes 6-lever Andersen Sampling apparatus |
LevelⅠ>7μm; Level Ⅱ: (4.7~7)μm; Level Ⅲ: (3.3~4.7)μm; Level Ⅳ: (2.1~3.3)μm; Level Ⅴ: (1.1~2.1)μm; Level Ⅵ: (0.6~1.1)μm |
Size of Aerosol Chamber
|
600×80×5mm (Length×Diameter×Thickness) |
Total number of positive quality control sampler particles
|
(2200±500)cfu |
Negative Chamber Flow Speed |
≥5m3/min |
Internal dimensions of operation Chamber |
740×940×440mm (L×W×H) |
Size of Main Machine |
1200×650×2100mm (L×W×H) |
Power Supply |
AC220V±10%, 50Hz |
Noise of the Tester |
<65dB(A) |
Weight
|
≈200kg |
Power Consumption
|
<1500W |
Accessories:
Standard Accessories |
1set |
Air compressor |
12pcs |
Petri dish |
|
1set |
Test tube |
|
3m |
Air pipe |
|
1pc |
Measuring cup |
|
1pc |
U disk |