Application: It is applicable to clothing with liquid-tight connections between different parts of the clothing and, if applicable, between the clothing and other items of personal protective equipment worn with it. |
Model |
GT-RC04 |
Diameter of the opening of the nozzle |
(4 ± 0,1) mm |
Length of the opening of the nozzle |
(4 ± 0,1) mm |
Distance between centre of spray rod to the center of turn table |
1500mm |
Inner diameter of the tube |
(12,5 ± 1) mm |
Distance between the opening of the nozzle and the manometer |
(80 ± 1) mm |
Standard |
ISO 17491-3 |
PS: A pressure gauge capable of providing an accurate reading at a liquid pressure of (150 ± 15) kPa shall be attached to the nozzle.