Programmable Temperature And Humidity Test Chamber 304 stainless steel material, can custom-made according different request.
Brand:
GESTERItem NO.:
GT-C52Order(MOQ):
1Product Detail
Programmable Temperature And Humidity Test Chamber GT-C52
Application of Environmental Test Chamber: Temperature And Humidity Test Chamber for conditioning of samples prior to testing. It also can be used for a variety of materials of high - low temperature alternating test. The test temperature, humidity and time can be programmed. |
1. Touch screen, South Korean import,900 temperature controller
2. Control mode: Touchpad
3. Temperature and Humidity Control Chamber Display: LCD display, Korean TEMi880, programmable.
4. System capacity:120 groups program X100 sections, required segmentation sections of each group can be arbitrarily divided and each program can freely link another.
5. Humidity And Temperature Controlled Chamber Cycle setting: each operation procedure can carry out 9999 times or heaps of times cycle and repeated performance, and to the degree, divide further 5 segmentation sections to do part cycle of the extra.
Capacity |
80L |
150L |
225L |
408L |
612L |
1000L |
Inside dimension WxHxD(cm) |
40x50x40 |
50x60x50 |
50x75x60 |
60x85x80 |
80x90x85 |
100X100X100 |
Outside dimension WxHxD(cm) |
88x136x99.5 |
100X146X110 |
123x136X120 |
110x171x137 |
130x181x137 |
152x 210x 127 |
Temperature range (optional, can be custom-made) |
-20~150℃ |
-40~150℃ |
-60~150℃ |
||
Humidity range |
20%~98%, can be set. |
||||
|
|||||
Rate of temperature increasing |
Room temperature~150℃ |
About 45 minutes(non-linear no-load, about 3.5℃/minute |
|||
Room temperature~-20℃ |
About 45 minutes(non-linear no-load, about 1.2℃/minute) |
||||
Machine precision |
Control accuracy (test precision) |
Temperature |
+/-0.5℃ |
||
Humidity |
+/-1%RH |
||||
Distribution accuracy (uniformity) |
Temperature |
+/-1.0℃ |
|||
Humidity |
+/-2%RH |
||||
Environment temperature |
5℃-28℃ |
||||
Environment relative humidity |
≤85%RH |
||||
Power voltage |
220V AC(+/-10%), protect ground wire, ground resistance≤4Ω. |
||||
Frequency |
50+/-0.5Hz |
||||
Power switch |
Air-break switch(40A, knife switch or outlet is prohibited to use.) |
||||
Power |
4.5KW |
||||
Humidification water |
Resistivity(≥500MΩ) |
||||
Requirement of ambient air quality |
Do not contain high concentration dust and flammable gas, without strong electromagnetic radiation source nearby. |
||||
Power supply |
AC1¢3W 380V 50/60HZ, voltage change allowed value: ±10% of fixed voltage. |
||||
Max loading capacity |
8KVA |
||||
Max operating current |
30A |
Related Tags :